Several standards are relevant to steam sterilizers used for medical purposes. They address different sizes of sterilizer as well as the sterilization process itself. The standard applicable in a specific case depends on the sterilizer and its intended use.
The most important standards include:
EN 285:2015+A1:2021
EN 285 specifies requirements and test methods for large steam sterilizers. It applies primarily to sterilizers used in healthcare facilities for the sterilization of medical devices and their accessories.
EN 13060:2025
EN 13060 specifies requirements and test methods for automatically controlled small steam sterilizers. It applies primarily to sterilizers for medical devices with a chamber volume of less than 60 l that are unable to accommodate a sterilization module of 300 mm × 300 mm × 600 mm. Small steam sterilizers used to sterilize liquids or pharmaceutical products are outside its scope.
EN ISO 17665:2024
EN ISO 17665 addresses the development, validation and routine control of moist-heat sterilization processes for medical devices. Its primary focus is control of the sterilization process rather than construction of the sterilizer itself.
The standards therefore complement each other: EN 285 and EN 13060 deal primarily with the technical performance of sterilizers, while EN ISO 17665 focuses on establishing and controlling a validated sterilization process.
During steam generation, most non-volatile water constituents do not evaporate and therefore become concentrated in the steam generator. Insufficient water quality can therefore cause deposits, corrosion and operational problems. Under unfavourable operating conditions, however, water droplets and dissolved or suspended substances can be entrained into the steam (carry-over).
Certain contaminants can therefore affect the chemical quality of the generated steam and, particularly through droplet entrainment, may reach the sterilized load.
For this reason, informative Annex B of EN 285 provides suggested maximum values for contaminants in feed water supplied to a dedicated steam generator.
The suggested maximum values for feed water supplied to a dedicated steam generator include:
It is important to note that Annex B of EN 285 is informative. The values are therefore suggested maximum values rather than a universally applicable feed-water specification for every sterilizer. The sterilizer manufacturer's requirements and any project-specific or regulatory requirements must also be considered.
For small steam sterilizers, EN 13060:2025 also contains water-related requirements and, in informative Annex F, suggested maximum limits of contaminants and specifications for water for steam sterilization. The applicable standard should therefore be established for the specific sterilizer before the water-treatment design is defined.
As the salt content increases, the amount of non-volatile material that can accumulate inside the steam generator also increases. Electrical conductivity therefore serves as an important summary parameter for dissolved ionic substances.
Low hardness also reduces the risk of mineral deposits. Calcium and magnesium compounds in particular can form deposits when water is heated and concentrated.
Such deposits can reduce heat-transfer efficiency, increase energy consumption and impair the operational reliability of the steam generator.
Chloride is a particularly important parameter for metallic materials at elevated temperatures. Increased chloride concentrations can promote corrosion processes and, under suitable conditions, contribute to localized corrosion of stainless steels.
For this reason, chloride is considered separately in the recommendations of EN 285.
Low conductivity alone is not always sufficient to fully assess water quality. Relevant individual water constituents should therefore also be considered when designing a water-treatment system.
Dissolved silicon species, listed as silicate in EN 285, can form poorly soluble deposits when water is thermally concentrated and suitable conditions are present. Such silicate-containing deposits can be particularly difficult to remove.
Where high feed-water and steam-quality requirements apply, silicate should therefore be considered during water-treatment design in addition to hardness, conductivity and chloride.
The necessary treatment process depends mainly on the raw-water composition and the requirements of the individual sterilizer.
Reverse osmosis is frequently used to produce low-salt feed water. Depending on the raw-water quality, pretreatment may include softening, filtration or other treatment stages.
Additional polishing processes may be required where particularly stringent water-quality requirements apply.
The objective is a consistently reproducible water quality with low hardness, low conductivity and low concentrations of critical contaminants.
The sterilizing effect is produced by steam. For steam to act reliably on the load, both its physical properties and its chemical quality need to be controlled. Water treatment is therefore an important part of the utility supply for a steam sterilizer. A properly designed treatment system helps to
Water quality should not be considered in isolation. Raw-water composition, water treatment, steam generation, the sterilizer and the sterilization process together form one technical system.